Diabetes Tech Lab

CGM MARD Comparison

We aggregate clinical trial data to compare the accuracy of continuous glucose monitors. Numbers reflect the overall MARD reported in pivotal FDA trials.

Device Overall MARD Wear Time Calibration
FreeStyle Libre 3 7.9% 14 Days Factory
Dexcom G7 8.2% 10 Days Factory / Optional
Eversense E3 8.5% 180 Days 1-2 per day
Dexcom G6 9.0% 10 Days Factory / Optional
FreeStyle Libre 2 9.2% 14 Days Factory
Medtronic Guardian 4 10.6% 7 Days Factory

Understanding the Data

MARD is calculated by comparing CGM readings to a lab-quality reference (usually YSI) during clinical trials. A lower number indicates the sensor readings align more closely with actual blood glucose levels.

While a sub-8% MARD is excellent, clinical MARD does not always translate directly to personal experience. Sensor placement, body chemistry, hydration, and rapidly changing glucose levels all impact real-world performance.

The "Day 1" Problem

Almost all sensors exhibit higher MARD (lower accuracy) during the first 12-24 hours of wear due to local tissue trauma from insertion. For example, Dexcom G7 reports a Day 1 MARD of 9.3%, which improves on subsequent days.